As the autonomously operating low temperature containers are air-cooled, only one power connection is required for operation.
In the event of a power blackout, the optional emergency power supply offers maximum safety. Redundant systems are used as standard to ensure that the sensitive goods are not exposed to any danger. For (off-grid) quality assurance, the temperature can be monitored separately by the customer. Lockable feed-throughs for the PI probes cable are available. Optional access control and/or video surveillance is also possible.
Benefits
- Airlock -20°C
- Fully redundant system
- Monitoring
- GMP qualified (DQ, IQ, OQ, PQ)
- TÜV certified
- available 24/7
optional
- Emergency generator
- Pre-freezer
- GPS tracking
Data Sheet
Cryogenic Container
Technical Data
Storage room
Cooling capacity: approx. 7.9 kW
Variable capacity control
Room temperature at operating point: -80°C
Technical Data
Airlock
Cooling capacity: approx. 3.3 kW
variable capacity control
Room temperature at operating point: -20°C
Container construction
Access
cryogenic cell: front double-wing door
Entrance / airlock: -20°C
Storage room: -80°C
Operating environment
up to 42°C outside temperature
humidity: max 60%
Electrical data
Supply voltage: 400V/50Hz/3Ph
Current consumption at nominal value: 50.7 A
Max. Operating current: 130 A
Power consumption at nominal value: 25kW
Special equipment
redundant switch cabinet design
redundant refrigerator
Cryogenic storage cell
Length: 5.35m
Width: 2.3 m
Height: 2.0m
Container Dimensions and weight
Length: 12.2 m
Width: 3.0 m
Height: 2.85 m
Weight: 11.000kg
Flexible customisation options
The standard version already offers the most important features. Should you require customised equipment, please do not hesitate to contact us.
Storage
options for
- Research
- Pharmaceutical industry
- Medical industry
- Vaccines